FDA Adverse Event Malfunction Summary report: N

INSIGNIA

MDR report key: 3241418 · Received July 22, 2013

Report

Report Number
2124215-2013-11705
Event Type
Malfunction
Date Received
July 22, 2013
Date of Event
April 4, 2013
Report Date
April 24, 2015
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

(B)(4) UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, TECHNICIANS USED AN ENGINEERING-LEVEL LONGEVITY PREDICTION CALCULATION TO ASSESS THE RATE OF BATTERY DEPLETION, GIVEN THE PROGRAMMED PARAMETERS AND OTHER DATA STORED WITHIN THE MEMORY OF THE DEVICE. THE RESULTS OF THIS CALCULATION INDICATED THAT THE ACTUAL RATE OF BATTERY DEPLETION FELL WITHIN AN ACCEPTABLE RANGE. NEXT, A COMPREHENSIVE SERIES OF DIAGNOSTIC TESTS WERE CONDUCTED THAT VERIFIED THE PERFORMANCE OF THERAPY DELIVERY, SENSING, AND RECORDING FUNCTIONS. HAVING MET THE ENGINEERING LONGEVITY PREDICTION, FUNCTIONALLY PASSING ALL RETURNED PRODUCT TESTING, AND WITH NO FURTHER INFORMATION TO INDICATE A PRODUCT PERFORMANCE ISSUE, WE HAVE CONCLUDED THAT THIS DEVICE EXPERIENCED NORMAL BATTERY DEPLETION.

Additional Manufacturer Narrative · 1

RECORDS INDICATE THIS DEVICE REMAINS IN SERVICE. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PACEMAKER DISPLAYED A REMAINING LONGEVITY OF 2.5 YEARS, HOWEVER THE MAGNET RATE WAS 90 PULSES PER MINUTE. TECHNICAL SERVICES DISCUSSED A MINOR CHANGE IN PACING PERCENTAGE, OUTPUTS OR IMPEDANCE MEASUREMENTS COULD CAUSE THE PROGRAMMER TO PREDICT A LONGER LIFE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

--

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340800 INSIGNIA IMPLANTABLE PULSE GENERATOR LWP GUIDANT CRM CLONMEL IRELAND 1290

Patients

Seq Age Sex Outcome Treatment
1 39 YR 1290| 4470| 4469