HOMECHOICE
Report
- Report Number
- 1416980-2013-19328
- Event Type
- Malfunction
- Date Received
- July 22, 2013
- Date of Event
- April 29, 2013
- Report Date
- June 28, 2013
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Removal / Correction Number
- 1423500-01/08/10-001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VE
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED FOR EVALUATION. REVIEW OF THE EVENT LOG IDENTIFIED THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT. THE CAUSE WAS DETERMINED TO BE ONE OR MORE CYCLES ADVANCED TO NEXT FILL WHEN SLOW OR NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2013 AT 22:48:20. DURING NIGHT DRAIN CYCLE FIVE, THE PATIENT'S ULTRAFILTRATION READING WAS 1422ML, INDICATING THE HOME PATIENT (HP) DRAINED 1422ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 341249 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | PD DISPOSABLES |