TI MATRIXMIDFACE SCREW SELF-DRILLING 6MM
Report
- Report Number
- 2520274-2013-04639
- Event Type
- Injury
- Date Received
- July 22, 2013
- Date of Event
- July 1, 2013
- Report Date
- July 2, 2013
- Manufacturer
- SYNTHES USA
- Product Code
- JEY
- PMA / PMN Number
- K083388
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE IS USED FOR TREATMENT, NOT DIAGNOSIS. PATIENT INFORMATION WAS REQUESTED BUT WAS NOT PROVIDED. INVESTIGATION COULD NOT BE COMPLETED AND NO CONCLUSION COULD BE DRAWN AS NO DEVICE WAS RETURNED AND NO LOT NUMBER WAS PROVIDED. PLACEHOLDER.
ADVERSE EVENT DUE TO SURGICAL DELAY OF 30 MINUTES OR GREATER. THE QUANTITY OF COMPLAINT SCREW WAS ORIGINALLY REPORTED AS ONE BUT ADDITIONAL INFORMATION RECEIVED BY THE SURGEON ON (B)(6) 2013, REPORTED MULTIPLE SCREWS. THE EXACT QUANTITY OF COMPLAINT SCREWS AND PART NUMBERS WAS REQUESTED, BUT NOT PROVIDED BY THE REPORTER. PLACEHOLDER.
ADDITIONAL INFORMATION PROVIDED BY THE SURGEON ON (B)(6) 2013, CONFIRMED THE COMPLAINT EVENT OCCURRED ON (B)(6) 2013. SURGEON STATED DURING THE PROCEDURE, MULTIPLE 5MM AND 6MM MATRIX SELF-DRILLING SCREWS BROKE WHILE PERFORMING AN OPEN REDUCTION INTERNAL FIXATION PROCEDURE TO TREAT A ZYGOMATICOMAXILLARY FRACTURE. THE COMPLAINT EVENT RESULTED IN MULTIPLE REPOSITIONING OF THE PLATES AND EXTENDED SURGICAL TIME BY 30 MINUTES OR MORE. PRE-DRILLING WAS NOT PERFORMED. THE THREADED PORTIONS OF THE BROKEN SCREWS REMAIN IN THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THOUGH REQUESTED, THE EXACT NUMBER OF BROKEN SCREWS AND THE ASSOCIATED PART NUMBERS WERE NOT PROVIDED BY THE SURGEON. THIS REPORT IS FOR AN UNKNOWN QUANTITY OF TI MATRIXMIDFACE SCREWS SELF-DRILLING 6MM. THIS FOLLOW-UP #1 REPORT IS 3 OF 3 FOR COMPLAINT (B)(4).
IT WAS REPORTED DURING AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE, THREE 6MM MATRIX SELF DRILLING SCREW HEADS BROKE OFF DURING INSERTION. THE THREADED PORTIONS OF THE THREE SCREWS REMAINED IN THE PATIENT. THE PROCEDURE WAS DELAYED BY APPROXIMATELY 15 MINUTES DUE TO THE INCIDENT. THIS REPORT IS FOR ONE, TI MATRIXMIDFACE SCREW SELF-DRILLING 6MM. THIS IS REPORT 2 OF 3 FOR COMPLAINT (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 341351 | TI MATRIXMIDFACE SCREW SELF-DRILLING 6MM | JEY | SYNTHES USA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 26 YR | Required Intervention |