FDA Adverse Event Malfunction Summary report: N

ORTHO SUMMIT SAMPLE HANDLING SYSTEM

MDR report key: 3241200 · Received September 28, 2006

Report

Report Number
2250051-2006-00414
Event Type
Malfunction
Date Received
September 28, 2006
Date of Event
August 31, 2006
Report Date
September 27, 2006
Manufacturer
HAMILTON BONADUZ AG
Product Code
MMH
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT HAS BEEN INVESTIGATED. THE FIELD SERVICE ENGINEER (FSE) EVALUATED THE INSTRUMENT AND FOUND A DISPOSABLE TIP HITTING THE PLATE IN THE REPORTED PROBLEM AREA OF THE PIPETTE ASSEMBLY. THE TIP CLAMP WAS REPLACED. THE INSTRUMENT WAS CLEANED, INSPECTED, TESTED WITHOUT FURTHER PROBLEM, AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE ORTHO SUMMIT SAMPLE HANDLING SYSTEM INSTRUMENT REPORTEDLY DID NOT PIPETTE SAMPLE/REAGENT AND DID NOT GENERATE AN ERROR MESSAGE. NO DEATH OR SERIOUS INJURY WAS ASSOCIATED WITH THIS INCIDENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC. COMPLAINT NUMBER: (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ORTHO SUMMIT SAMPLE HANDLING SYSTEM SAMPLE PIPETTOR MMH HAMILTON BONADUZ AG 7003009

Patients

Seq Age Sex Outcome Treatment
1 Other