9800

MDR report key: 3241081 · Received July 18, 2013

Report

Report Number
1720753-2013-08318
Event Type
Malfunction
Date Received
July 18, 2013
Date of Event
July 10, 2013
Report Date
July 18, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE SYSTEM SOFTWARE WAS RELOADED AND A FILAMENT CALIBRATION WAS PERFORMED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A HIGH MA (MILLIAMPS) ERROR MESSAGE DURING A PROCEDURE. THERE IS A NO REPORT OF PATIENT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
335422 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800