FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3241078 · Received July 18, 2013

Report

Report Number
1720753-2013-08327
Event Type
Malfunction
Date Received
July 18, 2013
Date of Event
July 2, 2013
Report Date
July 18, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND REPLACED THE BATTERY CHARGER CIRCUIT BOARD. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED "CHARGING ERROR". THE FSE NOTED THE SYSTEM DISPLAYED A PRECHARGE ERROR. THIS ERROR WILL LIKELY PREVENT THE SYSTEM FROM BOOTING. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
335529 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1