SPECTRUM INFUSION PUMP

MDR report key: 3240948 · Received July 17, 2013

Report

Report Number
1314492-2013-01368
Event Type
Malfunction
Date Received
July 17, 2013
Date of Event
June 1, 2013
Report Date
June 18, 2013
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE IN QUESTION. THE EVALUATION CONFIRMED CRACKS IN THE UPSTREAM SENSOR TAIL WHICH IS A KNOWN CONTRIBUTOR TO CONSTANT UPSTREAM OCCLUSION ALARMS. THE UPSTREAM SENSOR WAS REPLACED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PUMP ALARMED FOR AN UPSTREAM OCCLUSION WHEN NO OCCLUSION WAS PRESENT (MEDICATION, PROGRAMMED AMOUNT AND DELIVERY RATE UNKNOWN). IT WAS ALSO REPORTED THAT THERE WAS NO PATIENT INJURY.