SPECTRUM INFUSION PUMP
MDR report key: 3240948
·
Received July 17, 2013
Report
- Report Number
- 1314492-2013-01368
- Event Type
- Malfunction
- Date Received
- July 17, 2013
- Date of Event
- June 1, 2013
- Report Date
- June 18, 2013
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE IN QUESTION. THE EVALUATION CONFIRMED CRACKS IN THE UPSTREAM SENSOR TAIL WHICH IS A KNOWN CONTRIBUTOR TO CONSTANT UPSTREAM OCCLUSION ALARMS. THE UPSTREAM SENSOR WAS REPLACED.
Description of Event or Problem · 1
IT WAS REPORTED THAT A PUMP ALARMED FOR AN UPSTREAM OCCLUSION WHEN NO OCCLUSION WAS PRESENT (MEDICATION, PROGRAMMED AMOUNT AND DELIVERY RATE UNKNOWN). IT WAS ALSO REPORTED THAT THERE WAS NO PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 331966 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORPORATION | 35700 |