FDA Adverse Event Malfunction Summary report: N

LIBERTY CYCLER SET, SINGLE CONN./EXT. DL

MDR report key: 3240906 · Received July 17, 2013

Report

Report Number
8030665-2013-00457
Event Type
Malfunction
Date Received
July 17, 2013
Date of Event
June 21, 2013
Report Date
June 21, 2013
Manufacturer
REYNOSA MANUFACTURING
Product Code
FKX
PMA / PMN Number
K043363
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR PHYSICAL EVALUATION AND THE FAILURE MODE CANNOT BE CONFIRMED. HOWEVER, AN INVESTIGATION OF THE DEVICE MANUFACTURING RECORDS WAS CONDUCTED BY THE MANUFACTURER. THERE WERE NO DEVIATIONS OR NONCONFORMANCES DURING THE MANUFACTURING PROCESS. IN ADDITION, THE BATCH RECORD REVIEW CONFIRMED THE LABELING, MATERIAL, AND PROCESS CONTROLS WERE WITHIN SPECIFICATION.

Description of Event or Problem · 1

A PERITONEAL DIALYSIS PATIENT HAS REPORTED THAT DIALYSIS SOLUTION WAS LEAKING OUT OF THE CASSETTE AND INTO THE CYCLER. PATIENT WAS IN DRAIN FOUR OF TREATMENT. UPON REMOVING THE TUBING SET, SOLUTION WAS FOUND INSIDE THE CYCLER. THE ORIGIN OF THE LEAK COULD NOT BE IDENTIFIED. PATIENT DID NOT RECEIVE ANY PROPHYLACTIC ANTIBIOTICS AND HAS HAD NO ADVERSE EFFECTS. SAMPLE HAS NOT BEEN RETURNED TO THE MANUFACTURER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
331996 LIBERTY CYCLER SET, SINGLE CONN./EXT. DL FKX REYNOSA MANUFACTURING 13ER08800

Patients

Seq Age Sex Outcome Treatment
1 LIBERTY DIALYSIS CYCLER