FDA Adverse Event Malfunction Summary report: N

SINGLE USE SOFT-TIPPED DISPOSABLE INJECTOR

MDR report key: 3240877 · Received July 17, 2013

Report

Report Number
9611165-2013-00069
Event Type
Malfunction
Date Received
July 17, 2013
Date of Event
June 13, 2013
Report Date
June 20, 2013
Manufacturer
RAYNER INTRAOCULAR LENSES LTD.
Product Code
MSS
PMA / PMN Number
K062612
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE EVENT DESCRIPTION RECEIVED FROM THE HEALTHCARE FACILITY STATES THAT THE DOCTOR EXPERIENCED DIFFICULTIES WITH THE PLUNGER DURING USE; HOWEVER, PROCEEDED TO CONTINUE INJECTION. THE RAYNER IFU INCLUDES THE FOLLOWING STATEMENTS "WHEN DEPRESSING THE PLUNGER TOO MUCH RESISTANCE COULD INDICATE A TRAPPED LENS" AND "IF THE IOL CAUSES A BLOCKAGE IN THE INJECTOR SYSTEM, DISCARD THE INJECTOR". OUR REVIEW OF PRODUCTION RECORDS FOR THE R-INJ-04 INJECTOR BATCH B294 SHOWED THAT ALL MFG AND QUALITY CHECKS WERE CONDUCTED WITH SUCCESSFUL RESULTS. ALL INJECTORS RELEASED FOR DISTRIBUTION FROM THIS BATCH WERE WITHIN TOLERANCE, MET SPECIFICATION CRITERIA AND WERE WITHOUT DEFECTS. A REVIEW OF EXISTING VIGILANCE DATA FROM THE MONTH OF MFR OF THE R-INJ-04 INJECTOR (MAY 2012) WAS CONDUCTED IN ORDER TO DETERMINE IF ANY TRENDS EXISTED. THIS REVIEW CONCLUDED THAT NO OTHER INCIDENTS, OF ANY TYPE, HAVE BEEN RECEIVED AGAINST THE R-INJ-04 INJECTOR BATCH B294. THE RESULTS OF OUR DEVICE ANALYSIS/INSPECTION WILL BE PROVIDED IN A F/U REPORT.

Description of Event or Problem · 1

RAYNER INTRAOCULAR LENSES LIMITED RECEIVED NOTIFICATION OF AN EVENT THAT OCCURRED DURING USE OF A SINGLE USE SOFT-TIPPED DISPOSABLE INJECTOR (MODEL R-INJ-04). THE EVENT DESCRIPTION PROVIDED STATES "DURING INJECTION THE DOCTOR EXPERIENCED DIFFICULTIES WITH THE PLUNGER. THE DOCTOR OBSERVED THE IOL "INSIDE THE INJECTOR PLUNGER" AND IDENTIFIED THAT IT WAS "FRAGMENTED AND SEPARATED IN FRAGMENTS"".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
332795 SINGLE USE SOFT-TIPPED DISPOSABLE INJECTOR MSS - DISPOSABLE INJECTOR MSS RAYNER INTRAOCULAR LENSES LTD. R-INJ-04 B294

Patients

Seq Age Sex Outcome Treatment
1