FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 3240876
·
Received July 17, 2013
Report
- Report Number
- 1314492-2013-01345
- Event Type
- Malfunction
- Date Received
- July 17, 2013
- Date of Event
- June 1, 2013
- Report Date
- June 19, 2013
- Manufacturer
- BAXTER HEALTHCARE CORP
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY BAXTER. THIS COMPLAINT WAS OPENED DURING THE INVESTIGATION OF (B)(4). THE REPORTED "DOWNSTREAM OCCLUSION" ALARMS WERE CONFIRMED THROUGH THE EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED. IF ANY ADD'L INFO IS RECEIVED A F/U WILL BE SENT. THE DOWNSTREAM SENSOR AND DOWNSTREAM TUBING GUIDE WERE REPLACED.
Description of Event or Problem · 1
DURING THE EVAL OF A RETURNED SPECTRUM INFUSION PUMP, 150 OCCURRENCES OF A "DOWNSTREAM OCCLUSION" ALARM WERE IDENTIFIED IN THE EVENT HISTORY LOG. THERE WAS NO PT INJURY OR MEDICAL INTERVENTION REQUIRED SINCE THESE ITEMS WERE FOUND DURING EVAL OF THE DEVICE. NO ADD'L INFO IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 332105 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORP | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |