UV FLASH TRANSFER SETS
Report
- Report Number
- 1416980-2013-19284
- Event Type
- Malfunction
- Date Received
- July 22, 2013
- Date of Event
- June 28, 2013
- Report Date
- July 1, 2013
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- KDJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
(B)(4). THE DEVICE IS REPORTED TO BE AVAILABLE BUT HAS NOT BEEN RECEIVED. A REVIEW OF ALL BATCH RECORD DOCUMENTS FOR LOT NUMBER H13B06040 WAS PERFORMED WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION WERE NOTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
(B)(4). THE DEVICE WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION, LEAK TESTING, CLEAR PASSAGE TESTING, AND CLAMP FUNCTION TESTING WERE PERFORMED WITH NO ISSUES NOTED. THE INTEGRITY OF THE SEAL SURFACE WAS TESTED WITH NO LEAKS NOTED. THE INSIDE DIAMETER OF THE PATIENT CONNECTOR WAS MEASURED AND FOUND TO BE BELOW NOMINAL. IMPROVEMENTS WERE MADE TO THE MOLD AND THE MOLDING DEPARTMENT PROCEDURES TO ADDRESS THIS ISSUE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
THIS IS REPORT 3 OF 4 FOR THIS EVENT. IT WAS REPORTED THAT THE PATIENT CONNECTOR OF A UV FLASH TRANSFER SET THAT HAD A GAP BETWEEN THE TITANIUM ADAPTER AND THE TRANSFER SET WHEN IT HAD BEEN CHANGED OUT. THE TRANSFER SET WAS BEING USED WITH A PATIENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 340123 | UV FLASH TRANSFER SETS | SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE | KDJ | BAXTER HEALTHCARE - MOUNTAIN HOME | H13B06040 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | TITANIUM ADAPTER |