FDA Adverse Event Malfunction Summary report: N

NAVISTAR¿ DS ELECTROPHYSIOLOGY CATHETER

MDR report key: 3240477 · Received July 22, 2013

Report

Report Number
9673241-2013-00241
Event Type
Malfunction
Date Received
July 22, 2013
Date of Event
June 25, 2013
Report Date
June 26, 2013
Manufacturer
BIOSENSE WEBSTER, INC. (JUAREZ)
Product Code
OAD
PMA / PMN Number
P010068
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). IT WAS REPORTED THAT DURING AN ATRIAL FLUTTER (AFL) PROCEDURE WITH CARTO3, THE PHYSICIAN HEARD A POP AT THE END OF ABLATION. THE STOCKERT SETTINGS WERE TEMPERATURE LIMIT 65 C AND 50 WATT AND IT WAS TEMPERATURE CONTROL MODE. THE PHYSICIAN FOUND CHARRING AT THE TIP OF THE CATHETER. THERE WAS NO PATIENT CONSEQUENCE AND THE SIZE OF CHAR IS UNKNOWN. UPON RECEIPT, THE DEVICE WAS VISUALLY INSPECTED AND IT WAS FOUND IN NORMAL CONDITIONS. THEN PER THE REPORTED EVENT, THE CATHETER WAS TESTED AND IT PASSED ELECTRICAL, TEMPERATURE AND GENERATOR TESTS. THE DEVICE HISTORY RECORD WAS REVIEWED AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. THE CATHETER PASSED ALL SPECIFICATIONS. THE ROOT CAUSE OF THE POP AND CHAR REMAINS UNKNOWN.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN ATRIAL FLUTTER (AFL) PROCEDURE WITH CARTO3, THE PHYSICIAN HEARD A POP AT THE END OF ABLATION. THE STOCKERT SETTINGS WERE TEMPERATURE LIMIT 65 C AND 50 WATT AND IT WAS TEMPERATURE CONTROL MODE. THE PHYSICIAN FOUND CHARRING AT THE TIP OF THE CATHETER. THERE WAS NO PATIENT CONSEQUENCE AND THE SIZE OF CHAR IS UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
341307 NAVISTAR¿ DS ELECTROPHYSIOLOGY CATHETER CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER OAD BIOSENSE WEBSTER, INC. (JUAREZ) D-1201-17-S 15724307M

Patients

Seq Age Sex Outcome Treatment
1