FDA Adverse Event Malfunction Summary report: N

UV FLASH TRANSFER SETS

MDR report key: 3240463 · Received July 22, 2013

Report

Report Number
1416980-2013-19280
Event Type
Malfunction
Date Received
July 22, 2013
Date of Event
June 28, 2013
Report Date
July 1, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS REPORTED TO BE AVAILABLE BUT HAS NOT BEEN RECEIVED. A REVIEW OF ALL BATCH RECORD DOCUMENTS FOR LOT NUMBER H13A02017 WAS PERFORMED WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION WERE NOTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION, LEAK TESTING, CLEAR PASSAGE TESTING, AND CLAMP FUNCTION TESTING WERE PERFORMED WITH NO ISSUES NOTED. THE INTEGRITY OF THE SEAL SURFACE WAS TESTED WITH NO LEAKS NOTED. THE INSIDE DIAMETER OF THE PATIENT CONNECTOR WAS MEASURED AND FOUND TO BE BELOW NOMINAL. IMPROVEMENTS WERE MADE TO THE MOLD AND THE MOLDING DEPARTMENT PROCEDURES TO ADDRESS THIS ISSUE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

THIS IS REPORT 4 OF 4 FOR THIS EVENT. IT WAS REPORTED THAT THE PATIENT CONNECTOR OF A UV FLASH TRANSFER SET THAT HAD A GAP BETWEEN THE TITANIUM ADAPTER AND THE TRANSFER SET WHEN IT HAD BEEN CHANGED OUT. THE TRANSFER SET WAS BEING USED WITH A PATIENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340361 UV FLASH TRANSFER SETS SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME H13A02017

Patients

Seq Age Sex Outcome Treatment
1 TITANIUM ADAPTER