FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 3240389 · Received July 22, 2013

Report

Report Number
1416980-2013-19277
Event Type
Malfunction
Date Received
July 22, 2013
Date of Event
May 3, 2013
Report Date
June 27, 2013
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING THE EVENT HISTORY LOG REVIEW, AN INCREASED INTRA-PERITONEAL VOLUME EVENT WAS IDENTIFIED. THE ROOT CAUSE OF THE REPORTED ISSUE WAS DETERMINED TO BE USE ERROR, TIDAL TOTAL ULTRA FILTRATION REMOVAL SET TOO LOW. HOMECHOICE APD SYSTEMS TRAINER'S GUIDE GIVES INSTRUCTIONS ON HOW TO SET THE TIDAL THERAPY SETTINGS. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF ANY RELEVANT INFORMATION IS OBTAINED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2013 AT 20:48:20. DURING NIGHT DRAIN CYCLE SEVEN, PATIENT'S ULTRAFILTRATION READING WAS 1710 ML, INDICATING THE HOME PATIENT (HP) DRAINED 1310 ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000 ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
341169 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1