SUPERDIMENSION INREACH SYSTEM
Report
- Report Number
- 3004962788-2013-00023
- Event Type
- Injury
- Date Received
- July 18, 2013
- Date of Event
- June 21, 2013
- Report Date
- July 18, 2013
- Manufacturer
- SUPERDIMENSION INC.
- Product Code
- JAK
- PMA / PMN Number
- K092365
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE SITE RETURNED PROCEDURE RECORDING FOR EVALUATION. THE EVALUATION OF THE RECORDINGS FOUND THE SUPERDIMENSION SYSTEM WAS FINE BUT USER INTERFACE WHILE NAVIGATING CONTRIBUTED TO THE EVENT. ADDITIONALLY, THERE WAS A CT TO BODY DIVERGENCE AS THE CT WAS TAKEN IN FULL INHALATION, SO IT IS EXPECTED DURING THE PROCEDURE THE LUNGS WILL BE SHORTER WHICH RESULTED IN REGISTRATION ROTATIONAL ERROR. THE ERROR WAS SUCH THAT THE ANATOMY WAS SUPERIOR TO THE CT AND WHEN THE LOCATABLE GUIDE REACHED THE CT TARGET, IT WAS ACTUALLY MORE INFERIOR TO THE LESION. IT IS ESTIMATED THAT BECAUSE OF THIS WHEN TOOLS WERE INSERTED TO TAKE A BIOPSY THEY WERE GOING TOO INFERIOR CAUSING A PNEUMOTHORAX. A PNEUMOTHORAX IS A KNOWN SHORT TERM COMPLICATION WHEN A LUNG BIOPSY IS PERFORMED DURING A TRANSBRONCHIAL LUNG BIOPSY OR A CT-GUIDED PERCUTANEOUS BIOPSY.
SITE REPORTED A PNEUMOTHORAX IN A PATIENT. THE SUPER DIMENSION CASE WAS COMPLETED ON (B)(6) 2013 AND FLUORO EXAM IN THE OPERATING ROOM AT THE CONCLUSION OF THE CASE DID NOT REVEAL A PNEUMOTHORAX. A CHEST X-RAY SEVERAL HOURS LATER DID REVEAL A PNEUMOTHORAX AND PATIENT WAS ASYMPTOMATIC. REPEAT CHEST X-RAY ABOUT 3 HOURS LATER SHOWED NO CHANGE IN PNEUMOTHORAX AND STILL ASYMPTOMATIC AND PATIENT WAS SENT HOME. TWELVE TO EIGHTEEN HOURS LATER, PATIENT NOTED SOME CHEST PAIN AND WENT TO ANOTHER HOSPITAL WHERE THEY CONCLUDED THE PNEUMOTHORAX WAS INCREASING AND PLACED A PLEURAL CATHETER. THE PATIENT STAYED OVERNIGHT ON (B)(6) 2013, PNEUMOTHORAX RESOLVED AND CATHETER WAS PULLED. THE PATIENT WAS SEEN IN THE CLINIC ON (B)(6) 2013 AND A CHEST X-RAY REVEALED NO PNEUMOTHORAX.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 333325 | SUPERDIMENSION INREACH SYSTEM | ELECTROMAGNETIC NAVIGATION BRONCHOSCOPY | JAK | SUPERDIMENSION INC. | AAS00161-12C |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |