FDA Adverse Event Malfunction Summary report: N

TAXUS? LIBERTÉ?

MDR report key: 3240298 · Received July 22, 2013

Report

Report Number
2134265-2013-04888
Event Type
Malfunction
Date Received
July 22, 2013
Date of Event
May 9, 2013
Report Date
June 24, 2013
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
P060008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE IS A COMBINATION PRODUCT. DEVICE EVALUATED BY MFR: VISUAL AND MICROSCOPIC EXAMINATION OF THE RETURNED DEVICE IDENTIFIED THAT THE STENT WAS DAMAGED. THE STENT WAS STRETCHED ALONG ITS LENGTH CAUSING SOME OF THE STRUTS TO BE RAISED UP. THE BALLOON AND THE TIP OF THE DEVICE WERE VISUALLY AND MICROSCOPICALLY EXAMINED AND NO ISSUES WERE NOTED WITH THEIR PROFILES THAT COULD HAVE POTENTIALLY CONTRIBUTED TO THE COMPLAINT INCIDENT. THE BALLOON WAS TIGHTLY WRAPPED AND WAS NOT SUBJECTED TO ANY POSITIVE PRESSURE. NO KINKS OR DAMAGE WAS NOTED ALONG THE SHAFT OF THE DEVICE. NO OTHER ISSUES WERE NOTED WITH THE DEVICE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

REPORTABLE BASED ON ANALYSIS COMPLETED ON (B)(4) 2013. IT WAS REPORTED THAT DURING A STENTING TREATMENT PROCEDURE, FAILURE TO CROSS THE LESION OCCURRED. THE TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY (LAD). THE LESION WAS PRE-DILATED WITH APEX DILATATION CATHETER @ 16ATM. A 32 X 2.25MM TAXUS LIBERTÉ MONORAIL DRUG ELUTING STENT SYSTEM WAS ADVANCED TO CROSS THE LESION. SEVERAL ATTEMPTS AT CROSSING THE LESION WERE UNSUCCESSFUL. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENTS¿ CONDITION IS LISTED AS STABLE. HOWEVER, RETURNED PRODUCT ANALYSIS REVEALED STENT DAMAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
341136 TAXUS? LIBERTÉ? CORONARY DRUG-ELUTING STENT NIQ BOSTON SCIENTIFIC - GALWAY H7493894032220 15300623

Patients

Seq Age Sex Outcome Treatment
1 86 YR