FDA Adverse Event Malfunction Summary report: N

INFUSOR

MDR report key: 3240208 · Received July 22, 2013

Report

Report Number
1416980-2013-19265
Event Type
Malfunction
Date Received
July 22, 2013
Date of Event
June 26, 2013
Report Date
June 27, 2013
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A BATCH REVIEW WAS CONDUCTED AND REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. DURING INITIAL EVALUATION, VISUAL INSPECTION OF THE SAMPLE NOTED A RUPTURED BLADDER. MARKINGS ON THE INTERIOR AND EXTERIOR SURFACE OF THE BLADDER WERE OBSERVED NEAR THE RUPTURE LINE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). BAXTER RECEIVED ONE SAMPLE FOR EVALUATION. VISUAL EXAMINATION OF THE SAMPLE CONFIRMED THE REPORTED CONDITION. THE RESERVOIR WAS MICROSCOPICALLY EXAMINED AND THE CAUSE COULD NOT BE DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN INFUSOR'S RESERVOIR RUPTURED DURING FILLING. THE DEVICE WAS BEING FILLED MANUALLY VIA SYRINGE. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
339639 INFUSOR PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 13A073

Patients

Seq Age Sex Outcome Treatment
1