FDA Adverse Event Injury Summary report: N

TVC INSIGHT CATHETER

MDR report key: 3239943 · Received July 17, 2013

Report

Report Number
3004722468-2013-00001
Event Type
Injury
Date Received
July 17, 2013
Date of Event
June 14, 2013
Report Date
July 15, 2013
Manufacturer
INFRAREDX, INC.
Product Code
DQO
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

RESULTS FROM THE INFRAREDX INVESTIGATION OF THE ANGIOGRAPHIC FILMS AS WELL AS DISCUSSIONS WITH THE ATTENDING PHYSICIAN SUGGEST THAT THE ROOT CAUSE OF THIS EVENT MOST MORE LIKELY RELATED TO THE FIRST INFLATABLE DEVICE. DURING THIS FIRST INFLATION, THE WIRE WAS RETRACTED SLIGHTLY CAUSING THE TIP OF THE WIRE TO RESIDE IN THE TIGHT BEND OF AN ARTERY THAT EXPERIENCES A LARGER DEGREE OF CONTRACTION DURING EACH CARDIAC CYCLE.

Description of Event or Problem · 1

A (B)(6) YEAR OLD MAN, PRESENTED TO HOSPITAL WITH CHEST PAIN AND UNDERWENT CARDIAC CATHETERIZATION FOR A LESION LOCATED IN THE PROXIMAL RAMUS INTERMEDIUS. DURING THE PROCEDURE, INTRAVASCULAR ULTRASOUND WAS PERFORMED USING THE TVC IMAGING SYSTEM. POST REMOVAL OF TVC SYSTEM, AN INFLATABLE DEVICE WAS USED ON THE LESION. UPON REMOVAL OF THE INFLATABLE DEVICE, IT WAS DETERMINED THAT THE PT HAD A DISSECTION IN THE DISTAL RAMUS INTERMEDIUS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
331197 TVC INSIGHT CATHETER NIR/IVUS CORONARY IMAGING SYSTEM DQO INFRAREDX, INC. NIRS-MC7-7--3.2F 304234

Patients

Seq Age Sex Outcome Treatment
1 58 YR Life Threatening| R