FDA Adverse Event
Injury
Summary report: N
ECMO PUMP
MDR report key: 32388
·
Received April 29, 1996
Report
- Report Number
- MW1008965
- Event Type
- Injury
- Date Received
- April 29, 1996
- Date of Event
- April 8, 1996
- Report Date
- April 15, 1996
- Manufacturer
- SORIN BIOMEDICAL, INC.
- Product Code
- DTZ
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- OH, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
ECMO PUMP STOPPED FUNCTIONING AFTER "SNAPPING" SOUND HEARD. MANUAL CRANKING WAS REQUIRED WHILE THE PUMP WAS CHANGED. SPLIT BELT CONFIRMED BY BIOMEDICAL ENGINEERING. SECOND SUCH INCIDENT OF BELT BREAKAGE SINCE 1/96. PT'S O2 SATURATION DECREASED, BUT RETURNED TO BASELINE QUICKLY AFTER GOING BACK ON ECMO. ABG'S DURING HAND CRANKING 28/47/72/24. ALL APPROPRIATE PREVENTIVE MAINTENANCE HAD BEEN DONE ON PUMP. BELT WITH NORMAL USAGE TIME, SO SHOULD NOT HAVE SPLIT. RECEIVED NO RECOMMENDATION OR REPORT OF EVALUATION ON PREVIOUS INCIDENT REPORTED 1/96.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ECMO PUMP | OXYGENATOR | DTZ | SORIN BIOMEDICAL, INC. | 10-10-00 | G |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 4 MO | Required Intervention |