FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 3234154 · Received July 18, 2013

Report

Report Number
0001811755-2013-01699
Event Type
Malfunction
Date Received
July 18, 2013
Date of Event
June 28, 2013
Report Date
July 2, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON DISASSEMBLY FOR VISUAL INSPECTION, CORROSION WAS FOUND ON THE MOTOR, THE ROTOR, THE DRIVESHAFT AND THE SPINDLE HOUSING. THE SPINDLE HOUSING AND MOTOR HOUSING WERE ALSO DAMAGED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CORE IMPACTION DRILL HEATED UP DURING A CASE. THERE WAS NO PATIENT OR USER INJURY REPORTED AND NO ADVERSE CONSEQUENCES ALLEGED WITH THIS EVENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A BACKUP DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
333898 CORE IMPACTION DRILL DRIVER, WIRE, AND BONE DRILL, MANUAL DZJ STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1