ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-11312
- Event Type
- Malfunction
- Date Received
- July 18, 2013
- Report Date
- July 9, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS AT THE TIME OF THIS REPORT. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/28/2013 WITH THE FOLLOWING FINDINGS: WHEN THE PUMP WAS BOOTED UP THE DISPLAY SCREEN WAS DIM AND DISCOLORED. WHEN A TEST SCREEN REPLACED THE ORIGINAL DISPLAY SCREEN THE CONTRAST AND COLOR RETURNED TO NORMAL. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE DISPLAY SCREEN ON THEIR DEVICE WAS A BURNT BROWNISH COLOR. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 333761 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 6 YR |