FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3234005 · Received July 18, 2013

Report

Report Number
1416980-2013-18865
Event Type
Malfunction
Date Received
July 18, 2013
Date of Event
June 24, 2013
Report Date
June 24, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE REPORTED ISSUE WAS CONFIRMED DURING THE EVENT HISTORY LOG REVIEW. VISUAL INSPECTION, LEAK TESTING, CLEAR PASSAGE TESTING AND CLAMP FUNCTION TESTING WERE PERFORMED WITH NO ISSUES NOTED. SIMULATED THERAPY WAS PERFORMED ON THE HOMECHOICE SUCCESSFULLY. ALL PRESSURES WERE FOUND TO BE WITHIN SPECIFICATIONS. DURING EVALUATION NO FAILURE OR MALFUNCTION WAS IDENTIFIED THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED PROBLEM. THE CAUSE WAS DETERMINED TO BE ONE OR MORE CYCLES ADVANCES TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS REPORTED TO BE AVAILABLE BUT HAS NOT YET BEEN RECEIVED. SHOULD THE DEVICE BE RECEIVED OR ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A HOME PATIENT (HP) RECEIVED A HIGH DRAIN ERROR 106 ALARM ON A HOMECHOICE (HC) MACHINE WHEN THE HC WAS TURNED ON. THIS ALARM INDICATES THE PATIENT DRAINED GREATER THAN 200% OF THE MAXIMUM PRESCRIBED CYCLE FILL VOLUME IN STANDARD MODE. THE HP HAD NOT STARTED THERAPY YET AND WAS ABOUT TO START THERAPY FOR THE NIGHT. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE ALARM AND ARRANGED TO SWAP THE HC. THE HP WAS TO USE MANUAL SUPPLIES FOR THE NIGHT. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED IN ASSOCIATION WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
333181 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1 36 YR