MICRO SAGITTAL SAW
Report
- Report Number
- 0001811755-2013-01683
- Event Type
- Malfunction
- Date Received
- July 18, 2013
- Date of Event
- June 29, 2013
- Report Date
- June 29, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- HBE
- PMA / PMN Number
- K032303
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
THE REPORTED EVENT OF A BIAS CURRENT ERROR WAS NOT DUPLICATED; HOWEVER, IT WAS CONFIRMED THROUGH INSPECTION THAT THE MOTOR ASSEMBLY INCLUDING THE TPS FLEX WAS CORRODED.
DEVICE NOT YET RECEIVED. EVALUATION IS ANTICIPATED UPON RECEIPT OF DEVICE AND A FOLLOW-UP WILL BE FILED.
THE MICRO SAGITTAL SAW WAS TESTED DURING ROUTINE SERVICING. UPON EVALUATION IT DISPLAYED A BIAS CURRENT MESSAGE WHEN CONNECTED TO THE CONSOLE, SIGNALING A CONDITION OCCURRED FROM WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES TO THE USER ALLEGED WITH THIS EVENT.
THE MICRO SAGITTAL SAW WAS TESTED DURING ROUTINE SERVICING. UPON EVALUATION IT DISPLAYED A BIAS CURRENT MESSAGE WHEN CONNECTED TO THE CONSOLE, SIGNALING A CONDITION OCCURRED FROM WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES TO THE USER ALLEGED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 334522 | MICRO SAGITTAL SAW | DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) | HBE | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |