FDA Adverse Event Malfunction Summary report: N

OPTIFLOW JUNIOR INTERFACE

MDR report key: 3233842 · Received July 18, 2013

Report

Report Number
9611451-2013-00539
Event Type
Malfunction
Date Received
July 18, 2013
Report Date
June 18, 2013
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
CAT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
RESPIRATORY THERAPIST

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE OPTIFLOW JUNIOR CANNULA IS DESIGNED SPECIFICALLY FOR THE DELICATE ANATOMICAL FEATURES AND FLOW REQUIREMENTS OF NEONATAL AND PAEDIATRIC PATIENTS. IT FEATURES ADHESIVE PADS (WIGGLEPADS) TO MAINTAIN CANNULA STABILITY ON THE PATIENT'S CHEEKS, SOFT-TOUCH NASAL PRONGS AND BREATHABLE KINK-PROOF AND CRUSH-RESISTANT FLEXIBLE TUBING. THE OPTIFLOW JUNIOR PROVIDES A REVOLUTIONARY STEP BETWEEN LOW-FLOW OXYGEN THERAPY AND CPAP. METHOD: THE COMPLAINT OPT314 JUNIOR CANNULA WAS RECEIVED AT FISHER & PAYKEL HEALTHCARE AND WAS VISUALLY INSPECTED. RESULTS: INSPECTION OF THE CANNULA REVEALED THAT THE LOOP PAD (WIGGLEPAD PADDING) HAD BECOME DETACHED FROM THE LEFT SIDE OF THE CANNULA AND WAS MISSING. ADDITIONALLY, THE CANNULA WAS BROKEN IN HALF, BETWEEN THE NASAL PRONGS. A LOT CHECK WAS NOT PERFORMED AS LOT INFORMATION WAS NOT PROVIDED. CONCLUSION: THE HOSPITAL HAD INFORMED US THAT THE CANNULA HAD BROKEN IN HALF "WHEN STRETCHING TO RE-ATTACH". WE CAN CONCLUDE THAT THIS DAMAGE WAS THUS THE RESULT OF UNDUE FORCE BEING PLACED ON THE NECESSARILY DELICATE MATERIAL OF THE NASAL PRONGS. THE LEFT LOOP PAD (WIGGLEPAD) WAS NOT PROVIDED FOR EVALUATION SO WE WERE UNABLE TO DETERMINE WHAT HAD CAUSED IT TO DETACH. LOOP PADS ARE BONDED TO THE INFANT OPTIFLOW CANNULA USING SPECIALISED PRIMER AND CYANOACRYLATE GLUE, IN ADDITION TO THE ADHESIVE ON THE LOOP PAD'S BACKING. THE BOND STRENGTH OF THE LOOP PADS TO THE CANNULA EXCEEDS THE BOND STRENGTH OF THE LOOP PADS TO THE HOOK ON THE WIGGLEPADS. FISHER & PAYKEL HEALTHCARE MAINTAINS A CLOSE SUPERVISION OF THIS PRODUCT AND ITS BEHAVIOUR IN THE FIELD IN ORDER TO GAIN INFORMATION. WE ARE CONTINUOUSLY WORKING TO IMPROVE THE OPTIFLOW JUNIOR RANGE OF CANNULAE BASED ON CUSTOMER FEEDBACK.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE FIELD REPRESENTATIVE THAT AN OPT314 JUNIOR CANNULA BROKE IN HALF WHILE BEING REMOVED FROM A PATIENT. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
335680 OPTIFLOW JUNIOR INTERFACE CAT CAT FISHER & PAYKEL HEALTHCARE LTD OPT314 UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1