FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 3233790 · Received July 18, 2013

Report

Report Number
2531779-2013-11294
Event Type
Malfunction
Date Received
July 18, 2013
Report Date
July 1, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION 09/20/2013 - DEVICE EVALUATION:THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS 08/30/2013 WITH THE FOLLOWING FINDINGS: DURING EVALUATION, THE AUDIO BOLUS BUTTON COVER WAS FOUND TO BE COMPLETELY DETACHED FROM THE PUMP. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (ABB TACTILE CNG PRIOR TO DMG) ISSUE. IT WAS REPORTED THAT THE AUDIO BOLUS BUTTON HAD AN INTERMITTENT RESPONSE AND OCCASIONALLY REQUIRED MULTIPLE BUTTON PRESSES TO ELICIT A RESPONSE. REPORTEDLY, THE AUDIO BOLUS BUTTON HAD FALLEN OFF THE PUMP. IT WAS DETERMINED THAT THE TACTILE CHANGES HAD OCCURRED PRIOR TO THE DAMAGE TO THE BUTTON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
334607 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 19 YR