FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 3233757 · Received July 18, 2013

Report

Report Number
0001811755-2013-01674
Event Type
Malfunction
Date Received
July 18, 2013
Date of Event
June 27, 2013
Report Date
June 27, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON DISASSEMBLY FOR VISUAL INSPECTION CORROSION WAS FOUND ON THE MOTOR PINS. A BEARING WAS BROKEN, THE ROTOR COUPLER WAS WORN AND THE MOTOR HOUSING WAS BENT.

Description of Event or Problem · 1

WHILE TESTING THE CORE IMPACTION DRILL AT THE MANUFACTURER IT WAS REPORTED THAT THE DEVICE HEATED UP. THERE WAS NO PATIENT INVOLVEMENT OR ADVERSE CONSEQUENCES REPORTED WITH THIS EVENT.