FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 3233516 · Received July 18, 2013

Report

Report Number
1416980-2013-18862
Event Type
Malfunction
Date Received
July 18, 2013
Date of Event
May 18, 2013
Report Date
June 24, 2013
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS DETERMINED TO BE A FALSE EMPTY DETECT AND USE ERROR WITH THE INITIAL DRAIN ALARM SETTING INAPPROPRIATELY PROGRAMMED. THE HOMECHOICE / HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE STATES IN THE WARNING: "SETTING THE I-DRAIN ALARM VOLUME TOO LOW OR OFF CAN RESULT IN AN INCOMPLETE INITIAL DRAIN FOLLOWED BY A FULL FILL. THIS CAN RESULT IN AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) SITUATION." A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING THE EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED, WHICH OCCURRED DURING THE THERAPY INITIATED ON (B)(6) 2013 AT 22:41:11. DURING NIGHT DRAIN CYCLE ONE, THE PATIENT'S ULTRAFILTRATION READING WAS 1479ML, INDICATING THE HOME PATIENT (HP) DRAINED 1479ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1700ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
333741 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1