FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3233471 · Received July 18, 2013

Report

Report Number
1416980-2013-18846
Event Type
Malfunction
Date Received
July 18, 2013
Date of Event
April 25, 2013
Report Date
June 24, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION WAS COMPLETED. THIS EVENT REPRESENTS AN ANCILLARY SERVICE EVENT. THE IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS DETERMINED TO BE USE ERROR: THE TIDAL TOTAL ULTRAFILTRATION REMOVAL WAS SET TOO LOW. HOMECHOICE APD SYSTEMS TRAINER'S GUIDE GIVES INSTRUCTIONS ON HOW TO SET THE TIDAL THERAPY SETTINGS. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2013 AT 20:10:57. DURING NIGHT DRAIN CYCLE FIVE, THE PATIENT'S ULTRAFILTRATION READING WAS 1915 ML, INDICATING THE HOME PATIENT (HP) DRAINED 1570 ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2300 ML. NO ADDITIONAL INFORMATION IS AVAILABLE.