FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 3233439
·
Received July 18, 2013
Report
- Report Number
- 2531779-2013-11264
- Event Type
- Malfunction
- Date Received
- July 18, 2013
- Report Date
- June 19, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE PUMP DISPLAY LENS WAS FOUND TO HAVE MULTIPLE SCRATCHES. THE PUMP POWERED ON AND BOOTED TO THE VERIFY SCREEN WITH A FADED AND DISCOLORED DISPLAY SCREEN. THE DISPLAY SCREEN WAS REPLACED WITH A TEST SCREEN AND THE DISPLAY RETURNED TO NORMAL. UNRELATED TO THE DISPLAY ISSUE, THE KEYPAD WAS FOUND TO BE TORN AT THE OK BUTTON; ALL OF THE KEYPAD BUTTONS WERE CONFIRMED TO BE RESPONSIVE.
Description of Event or Problem · 1
THE PUMP WAS RETURNED TO ANIMAS AND WAS INVESTIGATED BY PRODUCT ANALYSIS ON 06/19/2013. INVESTIGATION FOUND THAT THE DISPLAY SCREEN WAS FADED AND DISCOLORED. THIS REPORT IS MADE BASED ON THE RESULTS OF INVESTIGATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 334282 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 16 YR |