FDA Adverse Event Malfunction Summary report: N

RESERVOIR 3ML

MDR report key: 3233342 · Received July 18, 2013

Report

Report Number
3004209178-2013-95778
Event Type
Malfunction
Date Received
July 18, 2013
Date of Event
July 15, 2013
Report Date
July 15, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
FRN
PMA / PMN Number
K032005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT INSULIN FROM THE RESERVOIR LEAKED DURING NORMAL USE. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
334842 RESERVOIR 3ML RESERVOIR FRN MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-332A H8492449

Patients

Seq Age Sex Outcome Treatment
1