FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3233333 · Received July 18, 2013

Report

Report Number
3004209178-2013-95811
Event Type
Injury
Date Received
July 18, 2013
Date of Event
June 27, 2013
Report Date
June 28, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING INCLUDING THE DISPLACEMENT TEST. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. NO MOISTURE DAMAGE ON THE MOTOR ASSEMBLY OR ELECTRONIC ASSEMBLY NOTED DURING VISUAL INSPECTION.

Description of Event or Problem · 1

THE CUSTOMER REPORTED BEING HOSPITALIZED DUE TO DIABETES KETOACIDOSIS AND HIGH BLOOD GLUCOSE OVER 600MG/DL. THE CALLER STATED THAT THE DOCTOR TOLD HER NOT TO USE THE DEVICE BECAUSE THE INSULIN FROM THE RESERVOIR LEAKED INTO THE RESERVOIR COMPARTMENT, AND SHE DID NOT RECEIVE ANY INSULIN. THE CUSTOMER STATED THAT SHE BOTTOMED OUT, AND SHE WAS GIVEN PEANUT BUTTER, CRACKERS, AND JUICE TO TREAT HER GLUCOSE LEVEL. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
334841 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAH

Patients

Seq Age Sex Outcome Treatment
1 51 YR Hospitalization