OT ULTRAMINI METER
Report
- Report Number
- 3008382007-2013-20254
- Event Type
- Injury
- Date Received
- July 18, 2013
- Report Date
- July 9, 2013
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K061118
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- OTHER
Narratives
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.
FOLLOW-UP # 1 ¿ (08/06/2013). THE PATIENT¿S METER HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 7/26/2013 AND 7/31/2013, RESPECTIVELY, WITH THE FOLLOWING FINDINGS:THE METER PASSED ALL TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE REPRODUCED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
FOLLOW-UP # 2 ¿ (08/22/2013). THE PATIENT¿S METER HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 8/6/2013 AND 8/16/2013, RESPECTIVELY, WITH THE FOLLOWING FINDINGS:THE METER PASSED ALL TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE CONFIRMED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON (B)(6) 2013 THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) ALLEGING HER ONETOUCH ULTRAMINI METER WAS DISPLAYING INACCURATELY HIGH READINGS COMPARED TO HER FEELINGS OR NORMAL RESULTS AND COMPARED TO ANOTHER METER . THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE ADVOCATE (CCA) DOCUMENTATION. THE PATIENT REPORTED THE ALLEGED ISSUE FIRST OCCURRED ON (B)(6) 2013 IN THE AFTERNOON. THE PATIENT REPORTED USING SELF ADJUSTING INSULIN TO MANAGE HER DIABETES. THE PATIENT ALLEGED OBTAINING READINGS OF ¿200, MAYBE 291 MG/DL¿ AND ¿291 MG/DL¿ COMPARED TO ¿40¿S MG/DL¿ ON A FREESTYLE METER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THE RESULTS EXCEEDS THE EXPECTED VALUE OF (B)(4) WHEN OBTAINED WITHIN 30 MINUTES. THE PATIENT REPORTED MAKING NO CHANGES TO HER USUAL DIET, ACTIVITY LEVEL OR MEDICATIONS ON THE DAY OF THE ALLEGED ISSUE. THE PATIENT REPORTED 10-15 MINUTES AFTER THE ALLEGED ISSUE OCCURRED SHE DEVELOPED SYMPTOMS OF FEELING ¿SHAKY AND FLASHING LIGHTS.¿ THE PATIENT DENIED HAVING ANY TREATMENT IN RESPONSE TO THE ALLEGED SYMPTOMS. AT THE TIME OF TROUBLESHOOTING, THE CCA DETERMINED THE METER WAS IN THE CORRECT UNIT OF MEASUREMENT AT THE TIME OF TESTING AND THE TEST STRIPS WERE IN GOOD CONDITION. HOWEVER A CONTROL SOLUTION TEST WAS NOT RUN DUE TO THE PATIENT NOT HAVING ANY CONTROL SOLUTION AT THE TIME OF THE CALL. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. THIS COMPLAINT IS BEING REPORTED AS AN ADVERSE EVENT BECAUSE THE PATIENT CLAIMS DUE TO THE ALLEGED ISSUE, SHE DEVELOPED SYMPTOMS SUGGESTIVE OF HYPOGLYCEMIA. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 335920 | OT ULTRAMINI METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3366658 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | Life Threatening |