FDA Adverse Event
Injury
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 3233322
·
Received July 18, 2013
Report
- Report Number
- 3004209178-2013-95810
- Event Type
- Injury
- Date Received
- July 18, 2013
- Date of Event
- June 27, 2013
- Report Date
- June 28, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST AS A RESULT OF A PROTRUDED/LOOSE DRIVE SUPPORT DISK. FURTHER TESTING COULD NOT BE PERFORMED DUE TO THE ALARM.
Description of Event or Problem · 1
THE CUSTOMER REPORTED BEING HOSPITALIZED DUE TO DIABETES KETOACIDOSIS AND HIGH BLOOD GLUCOSE OF 321MG/DL. THE CUSTOMER STATED THAT THE INSULIN PUMP IS SHOWING THAT IT IS PUMPING INSULIN, BUT IT WAS NOT. THE CALLER EXPERIENCED NAUSEA, VOMITING, AND EXCESSIVE THIRST. TROUBLESHOOTING WAS PERFORMED, AND THE DRIVE SUPPORT CAP WAS STICKING OUT. ADVISED THE CALLER TO DISCONNECT FROM THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 333373 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 26 YR | Hospitalization |