FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3233277 · Received July 18, 2013

Report

Report Number
3004209178-2013-95793
Event Type
Malfunction
Date Received
July 18, 2013
Date of Event
June 25, 2013
Report Date
June 25, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP UNABLE TO PRIME DURING THE PRIME TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. UNABLE TO PERFORM THE EXCESSIVE NO DELIVERY ALARM DUE TO PRIME ANOMALY.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP WAS NOT GIVING HER THE CORRECT DOSAGE OF INSULIN. THE BLOOD GLUCOSE READING AT TIME OF CALL WAS 219MG/DL. TROUBLESHOOTING WAS PERFORMED, AND THE ALARM HISTORY REVEALED A NO DELIVERY ALARM. THE CALLER STATED WHILE ATTEMPTING TO REWIND THE PISTON DID STOP, AND THE CUSTOMER KEPT REWINDING THE PISTON DOWN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
335684 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAH

Patients

Seq Age Sex Outcome Treatment
1 11 YR