FDA Adverse Event
Malfunction
Summary report: N
FLOGARD
MDR report key: 3233174
·
Received July 18, 2013
Report
- Report Number
- 1416980-2013-18845
- Event Type
- Malfunction
- Date Received
- July 18, 2013
- Date of Event
- June 24, 2013
- Report Date
- June 24, 2013
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. THIS IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE ROOT CAUSE OF THE DOOR OPEN/CLOSE CLAMP ALARM WAS DETERMINED TO BE DAMAGE TO THE DOOR. TO CORRECT THE CONDITION, THE PUMP HEAD DOOR ASSEMBLY WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.
Description of Event or Problem · 1
DURING PRODUCT EVALUATION BY A SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE A DOOR OPEN/CLOSE CLAMP ALARM. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 333694 | FLOGARD | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |