FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 3233163 · Received July 18, 2013

Report

Report Number
1416980-2013-18844
Event Type
Malfunction
Date Received
July 18, 2013
Date of Event
June 24, 2013
Report Date
June 24, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. THIS IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE ROOT CAUSE OF THE DAMAGED FORCE SENSING RESISTORS (FSRS)/TUBE MISLOADING SENSORS IS UNKNOWN. TO CORRECT THE CONDITION, THE FSRS/TUBE MISLOADING SENSORS WERE REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.

Description of Event or Problem · 1

DURING PRODUCT EVALUATION BY A SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE DAMAGED FORCE SENSING RESISTORS (FSRS). NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
336237 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1