FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3232990 · Received July 18, 2013

Report

Report Number
2531779-2013-11228
Event Type
Malfunction
Date Received
July 18, 2013
Report Date
July 4, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4)

Additional Manufacturer Narrative · 1

FOLLOW-UP #2, DATE OF SUBMISSION 10/28/2013. DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/16/2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS OBSERVED TO BE FULLY INTACT WITH NO EVIDENCE OF PEELING OR DAMAGE. THE CONTRAST, UP, DOWN, AND OK BUTTONS RESPONDED TO TESTING. THE KEYPAD WAS REMOVED TO INVESTIGATE AND CONTAMINATION WAS OBSERVED UNDER THE CONTRAST, DOWN, AND UP CONTACT BUTTONS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THE REPORTER REPORTS BUTTONS ON THE KEYPAD ARE ALL NOT RESPONDING. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.