PK DISSECTING FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2013-02602
- Event Type
- Malfunction
- Date Received
- July 18, 2013
- Date of Event
- June 14, 2013
- Report Date
- July 12, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K061260
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL
- Reporter Occupation
- NURSE
Narratives
THE INSTRUMENT WAS RETURNED BUT IS PENDING EVALUATION. A FOLLOW-UP MDR WILL BE SUBMITTED ONCE THE FAILURE ANALYSIS IS COMPLETED. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE PITCH UP CABLE WAS FOUND TO BE BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. NO OTHER DAMAGE FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE BROKEN CABLE IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS REPORTED THAT DURING A DA VINCI SI HYSTERECTOMY PROCEDURE, THE CUSTOMER NOTED A BROKEN WIRE ON THE PK DISSECTING FORCEPS INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.