FDA Adverse Event Malfunction Summary report: N

PK DISSECTING FORCEPS INSTRUMENT

MDR report key: 3232985 · Received July 18, 2013

Report

Report Number
2955842-2013-02602
Event Type
Malfunction
Date Received
July 18, 2013
Date of Event
June 14, 2013
Report Date
July 12, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K061260
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED BUT IS PENDING EVALUATION. A FOLLOW-UP MDR WILL BE SUBMITTED ONCE THE FAILURE ANALYSIS IS COMPLETED. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE PITCH UP CABLE WAS FOUND TO BE BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. NO OTHER DAMAGE FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE BROKEN CABLE IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI HYSTERECTOMY PROCEDURE, THE CUSTOMER NOTED A BROKEN WIRE ON THE PK DISSECTING FORCEPS INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
334621 PK DISSECTING FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420227-03 M10130321 401

Patients

Seq Age Sex Outcome Treatment
1 41 YR DA VINCI SI SYSTEM, ESU, INSTRUMENTS & ACCESSORIE