ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-11200
- Event Type
- Malfunction
- Date Received
- July 18, 2013
- Report Date
- June 26, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1 DATE OF SUBMISSION 09/17/2013-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/27/2013 WITH THE FOLLOWING FINDINGS:THE PUMP POWERED ON AND THE DISPLAY WAS FOUND TO BE CLEAR AND LEGIBLE. THE COMPLAINT COULD NOT BE CONFIRMED OR DUPLICATED ON INVESTIGATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. EVALUATION FOUND THAT THE PUMP POWERS UP PROPERLY.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. IT WAS REPORTED THAT THE DISPLAY WAS EXTREMELY DIM AND GRADUALLY BECOMING DIFFICULT TO READ. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 333776 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 14 YR |