FREESTYLE FREEDOM LITE
Report
- Report Number
- 2954323-2013-00386
- Event Type
- Injury
- Date Received
- July 18, 2013
- Date of Event
- June 14, 2013
- Report Date
- June 20, 2013
- Product Code
- NBW
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE METER WAS RETURNED AND TESTED WITH RETAINED TEST STRIPS. THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION AND NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. THE COMPLAINT OF 'TEST DOES NOT START' WAS NOT OBSERVED DURING PRODUCT INVESTIGATION. NOTE: THE REPORTED MEDICAL EVENT IS RELATED TO TWO ADDITIONAL METERS (CEGR312-M3678 AND CEGR187-M1522) AND REPORTED IN TWO ADDITIONAL MDRS.
CALLER (CUSTOMER'S WIFE) REPORTED THAT FOR APPROXIMATELY FIVE WEEKS THE CUSTOMER WAS UNABLE TO TEST USING HIS THREE ADC BLOOD GLUCOSE METERS DUE TO THE TESTS NOT STARTING AFTER THE BLOOD SAMPLES WERE APPLIED. CALLER FURTHER REPORTED THE CUSTOMER HAD AN APPOINTMENT AT THE VA HOSPITAL AT 10:00AM ON (B)(6) 2013 AND HAD HIS BLOOD GLUCOSE CHECKED. CUSTOMER RECEIVED A CALL FROM THE VA HOSPITAL AT APPROXIMATELY 5:00PM AND IT WAS STATED "HIS BLOOD SUGAR (WAS) AT 564 MG/DL (OBTAINED ON UNSPECIFIED HCP METER) AND HIS KETONES (WERE) ALSO HIGH". CUSTOMER WAS TAKEN TO THE EMERGENCY ROOM, TREATED WITH AN UNSPECIFIED INTRAVENOUS TREATMENT AND A GLASS OF WATER, AND WAS REPORTEDLY "NOT RESPONSIVE TO THE INSTRUCTION". CALLER FURTHER REPORTED THE CUSTOMER EXPERIENCED UNSPECIFIED SYMPTOMS AND DENIED SELF-TREATMENT. AN ADDITIONAL READING OF 461 MG/DL WAS OBTAINED WHILE CUSTOMER WAS IN THE EMERGENCY ROOM AND IT WAS UNKNOWN BY THE CALLER WHAT, IF ANY, DIAGNOSIS WAS RECEIVED BY THE CUSTOMER. THERE WAS NO REPORT OF DEATH OR PERMANENT IMPAIRMENT ASSOCIATED WITH THIS CASE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 333987 | FREESTYLE FREEDOM LITE | BLOOD GLUCOSE MONITORING SYSTEM | NBW | 1352835 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR | Hospitalization| R |