FDA Adverse Event Injury Summary report: N

MEDOS* PROG INFANT VALVESYSTEM

MDR report key: 3232891 · Received July 18, 2013

Report

Report Number
1226348-2013-21481
Event Type
Injury
Date Received
July 18, 2013
Manufacturer
CODMAN & SHURTLEFF, INC.
Product Code
JXG
PMA / PMN Number
PK974739
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

HISTORICALLY, FOR COMPLAINTS OF THIS NATURE, EXAMINATIONS OF THE VALVES HAVE REVEALED THE ACCUMULATIONS OF BIOLOGICAL DEBRIS WITHIN THE VALVES, WHICH HAVE RESULTED IN BLOCKAGES WITHIN THE DEVICES. REVIEWS OF THE DEVICE HISTORY RECORDS HAVE CONFIRMED THAT THE VALVES HAVE MET SPECIFICATIONS WHEN RELEASED TO STOCK. A FOLLOW UP REPORT WILL BE FILED IF THE INVESTIGATION OF THIS DEVICE REVEALS A RESULT DIFFERENT FROM THAT WHICH IS STATED ABOVE OR IF ANY FURTHER INFORMATION REGARDING THE CAUSE OR MODE OF FAILURE IS MADE AVAILABLE. OTHERWISE, THE COMPLAINT IS CLOSED AT THIS TIME.

Description of Event or Problem · 1

THE AFFILIATE REPORTED THAT THE DEVICE WAS IMPLANTED VIA V-P SHUNT. AFTER IMPLANTATION IT WAS NOTED THAT THE VENTRICLES OF THE PATIENT¿S BRAIN BECAME LARGE. AS A RESULT, THE DEVICE WAS REVISED DUE TO BLOCKAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
335137 MEDOS* PROG INFANT VALVESYSTEM SHUNT, CENTRAL NERVOUS SYSTEM & COMPS JXG CODMAN & SHURTLEFF, INC.

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention