TALENT AAA
Report
- Report Number
- 2953200-2013-01382
- Event Type
- Injury
- Date Received
- July 18, 2013
- Date of Event
- June 21, 2013
- Report Date
- June 21, 2013
- Manufacturer
- MEDTRONIC IRELAND
- Product Code
- MIH
- PMA / PMN Number
- P070027
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). RESULTS: INHERENT RISK OF PROCEDURE (STENT GRAFT MIGRATION); PATIENT'S CONDITION AFFECTED EFFECTIVENESS OF DEVICE (DISEASE PROGRESSION, NECK DILATATION). CONCLUSION: DEVICE FAILURE/LACK OF EFFECTIVENESS RELATED TO PATIENT CONDITION (DISEASE PROGRESSION, NECK DILATATION).
A TALENT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. ANEURYSM AND VESSEL MORPHOLOGY AT THE TIME OF IMPLANT IS UNKNOWN. DURING A FOLLOW UP VISIT IT WAS DETERMINED THAT THE TALENT STENT GRAFT HAD MIGRATED DISTALLY DUE TO DISEASE PROGRESSION AND NECK DILATATION. THE PHYSICIAN FOUND NO SIGNS OF ENDOLEAK. AN ENDURANT CUFF WAS ADDED AND THE INTERVENTION WAS COMPLETED SUCCESSFULLY. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 334206 | TALENT AAA | SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT | MIH | MEDTRONIC IRELAND | V00413084 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00069 YR | Required Intervention |