FDA Adverse Event Injury Summary report: N

HAKIM PROGRAMMABLE VALVE WITH PRECHAMBER

MDR report key: 3232862 · Received July 18, 2013

Report

Report Number
1226348-2013-21482
Event Type
Injury
Date Received
July 18, 2013
Date of Event
June 25, 2013
Manufacturer
CODMAN & SHURTLEFF, INC.
Product Code
JXG
PMA / PMN Number
PK974739
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

HISTORICALLY, FOR COMPLAINTS OF THIS NATURE, EXAMINATIONS OF THE VALVES HAVE REVEALED THE PRESENCE OF BIOLOGICAL DEBRIS WITHIN THE DEVICES WHICH HAVE INTERFERED WITH THEIR PROGRAMMING AND/OR PRESSURE CONTROL MECHANISMS. THE BIOLOGICAL DEBRIS HAS CAUSED THE RETURNED VALVES TO FAIL THE PROGRAMMING AND/OR PRESSURE TESTS AND APPEARS TO HAVE CAUSED THE DIFFICULTIES EXPERIENCED BY THE CUSTOMERS. REVIEWS OF THE DEVICE HISTORY RECORDS HAVE CONFIRMED THAT THE VALVES HAVE MET SPECIFICATIONS WHEN RELEASED TO STOCK. A FOLLOW UP REPORT WILL BE FILED IF THE INVESTIGATION OF THIS DEVICE REVEALS A RESULT DIFFERENT FROM THAT WHICH IS STATED ABOVE OR IF ANY FURTHER INFORMATION REGARDING THE CAUSE OR MODE OF FAILURE IS MADE AVAILABLE. OTHERWISE, THE COMPLAINT IS CONSIDERED TO BE CLOSED AT THIS TIME.

Description of Event or Problem · 1

THE AFFILIATE REPORTED THAT AFTER IMPLANTATION OF THE DEVICE, IT WAS NOTED THAT THE DEVICE WOULD NOT REPROGRAM THE PRESSURE. THE PRESSURE COULD NOT BE CHANGED EVEN WITH NEODYMIUM. AS A RESULT, THE DEVICE WAS REVISED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
334197 HAKIM PROGRAMMABLE VALVE WITH PRECHAMBER SHUNT, CENTRAL NERVOUS SYSTEM & COMPS JXG CODMAN & SHURTLEFF, INC.

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention