MARYLAND BIPOLAR FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2013-02593
- Event Type
- Malfunction
- Date Received
- July 18, 2013
- Date of Event
- May 21, 2013
- Report Date
- June 19, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING EVALUATION FOUND THAT THE PITCH UP CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINED THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. THE OTHER CABLES AT THE WRIST WERE UNDAMAGED. ENGINEERING ALSO FOUND MAIN TUBE DAMAGE. THE DISTAL END OF THE MAIN TUBE HAD VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE .075 - .190 IN LENGTH AND NOT ALIGNED WITH THE MAIN TUBE AXIS. THE EVIDENCE WAS INCONCLUSIVE, BUT THE DAMAGE WAS LIKELY CAUSED BY MISHANDLING/MISUSE. NO OTHER DAMAGE WAS FOUND. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS - HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS REPORTED THAT DURING A DA VINCI SI PROSTATECTOMY PROCEDURE THE MARYLAND BIPOLAR FORCEPS INSTRUMENT CABLES WERE BROKEN NEAR THE TIP. NO MISSING OR FALLEN PIECES WERE REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 333184 | MARYLAND BIPOLAR FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420172-07 | M11130112 964 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DAVINCI SI SYSTEM, ESU, INSTRUMENTS & ACCESSORIES |