UNICEL® DXH 800 COULTER® CELLULAR ANALYSIS SYSTEM
Report
- Report Number
- 1061932-2013-01416
- Event Type
- Malfunction
- Date Received
- July 18, 2013
- Date of Event
- June 24, 2013
- Report Date
- June 25, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K081930
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE. THE FSE INSPECTED THE INSTRUMENT AND FOUND HEMOGLOBIN (HGB) RUNNING LOW AND OUT ON QC. THE FSE DETERMINED THAT HGB OUTPUT IS TOO LOW (WOULD NOT GO ABOVE 0.38 AND THE NORMAL VALUE IS 0.60). THE FSE REPLACED THE HGB PREAMPLIFIER, RESOLVING THE ISSUES. AS PART OF PREVENTATIVE MAINTENANCE, THE FSE ALSO INSTALLED MULTI-TRANSDUCER MODULE (MTM) OPTICAL COVERS AND REPLACED THE HGB LAMP. UPON FURTHER INSPECTION, THE FSE DISCOVERED THAT LMALS (LOWER MEDIAN ANGLE LIGHT SCATTER) OFFSET VOLTAGE WAS ELEVATED. THE FSE REPLACED THE MALS (MEDIAN ANGLE LIGHT SCATTER) SENSOR TO RESOLVE THE ISSUES. THE FSE ALSO REALIGNED THE MULTI-TRANSDUCER MODULE (MTM) AND MONITORED LATRON GAIN CALIBRATIONS AS PART OF INCIDENTAL MAINTENANCE, AND VERIFIED THE INSTRUMENT'S PERFORMANCE. ONE FAILURE MODE WAS RELATED TO THE HGB PREAMPLIFIER IN THE HGB ASSEMBLY AND ANOTHER FAILURE MODE WAS ATTRIBUTED TO THE MALS SENSOR. BECKMAN INTERNAL IDENTIFIER NUMBER FOR THIS REPORT IS (B)(4).
THE CUSTOMER REPORTED TO BECKMAN COULTER (BEC) THAT THE UNICEL DXH 800 COULTER ANALYZER DXH800 WAS RECOVERING LOW HEMOGLOBIN (HGB) VALUES ON QUALITY CONTROLS (QC) SAMPLES. THE CUSTOMER INDICATED THAT THERE WERE NO ERRONEOUS PATIENT RESULTS GENERATED OR REPORTED OUTSIDE OF THE LABORATORY AS THE ISSUES OCCURRED WHILE RUNNING QC. NO DEATH OR INJURY IS ATTRIBUTED TO THIS EVENT AND THERE WAS NO CHANGE TO PATIENT TREATMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 334270 | UNICEL® DXH 800 COULTER® CELLULAR ANALYSIS SYSTEM | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | DXH 800 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |