UNIVERSAL DRILL
Report
- Report Number
- 0001811755-2013-01667
- Event Type
- Malfunction
- Date Received
- July 18, 2013
- Date of Event
- June 24, 2013
- Report Date
- June 25, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
THE REPORTED EVENT OF THE DEVICE HAVING RUN ON WAS NOT CONFIRMED DURING FUNCTIONAL EVALUATION BY A MANUFACTURER TECHNICIAN. BASED ON A REVIEW OF THE RISK DOCUMENT, THE DEVICE CAN RUN WITHOUT ACTIVATION DUE TO MAGNETIC FIELD INTERFERENCE IN THE ENVIRONMENT OF THE DEVICE. A MAGNETIC FIELD CAN ACTIVATE THE DRILL'S HALL SENSOR IF IT IS TOO CLOSE TO THE DRILL, WHICH IS A POSSIBLE CAUSE OF THE REPORTED EVENT. DURING EVALUATION, CORROSION WAS FOUND THROUGHOUT SEVERAL INTERNAL COMPONENTS OF THE DEVICE. THE DEVICE WAS SERVICED AND RETURNED TO THE USER FACILITY.
FAILURE ANALYSIS IN PROGRESS. FOLLOW UP WILL BE SUBMITTED ONCE QUALITY INVESTIGATION IS COMPLETE.
IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED AT THE USER FACILITY THE DEVICE RAN WITHOUT USER ACTIVATION AFTER THE TRIGGER WAS RELEASED. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED AT THE USER FACILITY THE DEVICE RAN WITHOUT USER ACTIVATION AFTER THE TRIGGER WAS RELEASED. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 333619 | UNIVERSAL DRILL | DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE | ERL | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |