FDA Adverse Event Malfunction Summary report: N

UNIVERSAL DRILL

MDR report key: 3232566 · Received July 18, 2013

Report

Report Number
0001811755-2013-01667
Event Type
Malfunction
Date Received
July 18, 2013
Date of Event
June 24, 2013
Report Date
June 25, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED EVENT OF THE DEVICE HAVING RUN ON WAS NOT CONFIRMED DURING FUNCTIONAL EVALUATION BY A MANUFACTURER TECHNICIAN. BASED ON A REVIEW OF THE RISK DOCUMENT, THE DEVICE CAN RUN WITHOUT ACTIVATION DUE TO MAGNETIC FIELD INTERFERENCE IN THE ENVIRONMENT OF THE DEVICE. A MAGNETIC FIELD CAN ACTIVATE THE DRILL'S HALL SENSOR IF IT IS TOO CLOSE TO THE DRILL, WHICH IS A POSSIBLE CAUSE OF THE REPORTED EVENT. DURING EVALUATION, CORROSION WAS FOUND THROUGHOUT SEVERAL INTERNAL COMPONENTS OF THE DEVICE. THE DEVICE WAS SERVICED AND RETURNED TO THE USER FACILITY.

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS IN PROGRESS. FOLLOW UP WILL BE SUBMITTED ONCE QUALITY INVESTIGATION IS COMPLETE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED AT THE USER FACILITY THE DEVICE RAN WITHOUT USER ACTIVATION AFTER THE TRIGGER WAS RELEASED. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED AT THE USER FACILITY THE DEVICE RAN WITHOUT USER ACTIVATION AFTER THE TRIGGER WAS RELEASED. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
333619 UNIVERSAL DRILL DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE ERL STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1