ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-11146
- Event Type
- Malfunction
- Date Received
- July 18, 2013
- Report Date
- June 26, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1: DATE OF SUBMISSION 10/14/2013. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/26/2013 WITH THE FOLLOWING FINDINGS: THE CONTRAST, UP, AND DOWN BUTTONS WERE RESPONDING INTERMITTENTLY. THERE WAS NO DAMAGE TO THE KEYPAD. THE LEAK TEST FAILED DUE TO A CRACK IN THE PUMP AT THE RIGHT UPPER CORNER BETWEEN THE LENS AND THE CASE SEAL. THE KEYPAD WAS REMOVED AND CONTAMINATION WAS FOUND UNDER THE CONTRAST, UP, AND DOWN BUTTON CONTACTS. UNRELATED TO THE REPORTED ISSUE, CORROSION WAS FOUND ON THE FORCE SENSOR CIRCUITS, VIBRATOR MOTOR, THE BOTTOM OF THE CASE, AND BT1. MOISTURE WAS FOUND IN PUMP. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES W/MOISTURE) ISSUE. THE REPORTER ALLEGED THAT, AFTER SWIMMING, THE KEYPAD BUTTONS WERE NOT RESPONDING APPROPRIATELY AND FELT LIKE THERE WAS MOISTURE BENEATH THEM. IT WAS ALSO NOTED THAT OK BUTTON WAS STICKING. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 334910 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR |