FDA Adverse Event Malfunction Summary report: N

PRECISE BIPOLAR FORCEPS INSTRUMENT

MDR report key: 3232509 · Received July 18, 2013

Report

Report Number
2955842-2013-02586
Event Type
Malfunction
Date Received
July 18, 2013
Date of Event
June 14, 2013
Report Date
June 20, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED ON (B)(4) 2013. ENGINEERING NOTED BENT GRIPS AND MAIN TUBE DAMAGE WITH THE RETURNED INSTRUMENT. ONE GRIP WAS BENT, WHICH CAUSES SIDE TO SIDE MISALIGNMENT OF GRIPS. THERE WAS A 0.150 OFFSET AT THE TIPS. THE BENT GRIP HAD SEPARATION OF THE YAW PULLEY AND PULLEY COVER AT THE GLUE JOINT, INDICATING OVERLOADING AT THE TIP. THE DISTAL END OF MAIN TUBE HAD VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. SCRATCHES WERE 0.080 - 0.145 IN LENGTH AND WERE NOT ALIGNED WITH THE TUBE AXIS. NO OTHER DAMAGE WAS FOUND. EVIDENCE NOT CONCLUSIVE, BUT MAIN TUBE DAMAGE MAY HAVE BEEN CAUSED BY MISHANDLING/MISUSE. THE INSTRUMENTS & ACCESSORIES INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS -HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE MAIN TUBE DAMAGE FOUND DURING FAILURE ANALYSIS COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT, IF TO REOCCUR.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED INTUITIVE SURGICAL, INC. ABOUT THE PRECISE BIPOLAR FORCEPS INSTRUMENT BUT NO DESCRIPTION OF THE INSTRUMENT WAS PROVIDED. NOTHING REPORTEDLY FELL INTO A PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
336049 PRECISE BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420110-06 M10121016 813

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES