FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC ACE

MDR report key: 3232488 · Received July 18, 2013

Report

Report Number
3005075853-2013-03603
Event Type
Malfunction
Date Received
July 18, 2013
Date of Event
June 28, 2013
Report Date
July 1, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
LFL
PMA / PMN Number
K042777
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED WITH THE DISTAL TIP OF THE BLADE BROKEN OFF AND NOT RETURNED WITH THE DEVICE. THE REMAINING BLADE PORTION WAS SCRATCHED ¿ EVIDENCE OF CONTACT WITH METAL IN OR OUT OF THE OPERATIVE FIELD. THE DEVICE WAS ACTIVATED WITH THE GENERATOR AND AN ERROR CODE 5 WAS DISPLAYED.PROBABLE CAUSES OF BLADE DAMAGE, INCLUDING BREAKAGE, ARE EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE, BLADE CONTACT WITH OTHER DEVICES, STAPLES OR CLIPS DURING THE PROCEDURE OR USING ANY MEANS OTHER THAN THE BLADE WRENCH TO ATTACH OR DETACH THE BLADE. ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE DAMAGE SEVERITY AND RESULT IN AN ERROR CODE 5 OR BLADE ¿LOCKOUT¿ LATER IN THE PROCEDURE, AND CONTINUED USAGE CAN RESULT IN A BROKEN BLADE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC SIGMOIDECTOMY, AN ERROR 5 WAS DISPLAYED MAY TIMES DURING USE ON THE COLON. THE BLADE WAS BROKEN OFF WHEN THE DEVICE HAD BEEN OUTSIDE THE PATIENT. NO BROKEN PIECE FELL INTO THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ONE DEVICE WILL BE RETURNING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
336010 ULTRACISION HARMONIC ACE INSTRUMENT, ULTRASONIC SURGICAL LFL ETHICON ENDO-SURGERY, LLC. UNK J92J96

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR, HANDPIECE