MIS SINGLE INNER SETSCW
Report
- Report Number
- 1526439-2013-21468
- Event Type
- Malfunction
- Date Received
- July 18, 2013
- Date of Event
- May 6, 2013
- Report Date
- June 25, 2013
- Manufacturer
- DEPUY SYNTHES SPINE
- Product Code
- NKB
- PMA / PMN Number
- PK041801
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PHYSICIAN
Narratives
A FOLLOW UP REPORT WILL BE FILED UPON COMPLETION OF THE INVESTIGATION.
VISUAL EXAMINATION OF THE RETURNED SET SCREW CONFIRMED THE THREADS HAD TORN AWAY FROM THE DEVICE. A REVIEW OF THE DHR ACKNOWLEDGED THAT NO ISSUES WERE IDENTIFIED DURING THE MANUFACTURING AND RELEASE OF THIS PRODUCT THAT COULD HAVE BEEN ATTRIBUTED TO THE PROBLEM REPORTED BY THE CUSTOMER. THE PRODUCT WAS RELEASED ACCOMPLISHING ALL QUALITY REQUIREMENTS. NO DEFINITIVE CONCLUSIONS CAN BE DRAWN AS TO THE CAUSE THE TORN THREADS. HOWEVER, THE DAMAGE SUGGESTS THAT THE SET SCREW BECAME CROSS THREADED DURING ADVANCEMENT WITHIN THE CONCOMITANT DEVICE PEDICLE SCREW. IN THE ABSENCE OF AN IDENTIFIED DEVICE MANUFACTURING/RELEASE ISSUE OR OBSERVED TREND, THIS COMPLAINT WILL BE CLOSED WITH NO FURTHER ACTION REQUIRED.
IT WAS REPORTED THAT DURING AN MIS PROCEDURE, THE THREADS OF TWO SET SCREWS BECAME DEFORMED DURING ADVANCEMENT. AS A RESULT THE SET SCREWS COULD NOT BE ASSEMBLED TO MATING PEDICLE SCREW. ALSO, THE INNER THREADS OF THE MATING PEDICLE SCREW WERE ALSO DEFORMED. HOWEVER, RECEIPT OF THE COMPLAINT SAMPLES ON (B)(4) 2013 FOUND THREADS HAD BEEN TORN FROM THE SET SCREWS AND NOT SIMPLY DEFORMED AS REPORTED. CONCOMITANT DEVICE: PEDICLE SCREW, CATALOG NUMBER UNKNOWN. SEE MFG MEDWATCH REPORT NO. 1526439-2013-21467 FOR THE FIRST SET SCREW THAT WAS INVOLVED IN THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 333152 | MIS SINGLE INNER SETSCW | ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE | NKB | DEPUY SYNTHES SPINE | ALLCR4 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |